Installation Conditions For Infrared Carbon And Sulfur Analyzer

Infrared carbon and sulfur analyzer is a high precision measuring instrument. It consists of infrared carbon and sulfur detection part, high frequency induction combustion furnace, electronic balance, computer, printer and five parts. Analysis of carbon and sulfur content in ferrous metals, non-ferrous metals, rare earth metals, coal, coke, inorganics, ore, ceramics and other materials in industries such as coke, commodity inspection, scientific research, and chemical industry.



1. Infrared carbon and sulfur analyzer analysis laboratory requirements: Keep away from places with high humidity, dust, corrosive gases such as large earthquake sources, acid and alkali, and the area is larger than 3m * 3m. The indoor temperature should be 10-30 degrees Celsius air conditioning).
 
2. Infrared carbon and sulfur analyzer analysis laboratory power equipment requirements: voltage AC220V ± 5%, frequency 50Hz ± 2%, the diameter of the power supply wire introduced in the laboratory should be determined according to the total power of laboratory equipment, power supply is well grounded, experiments The grounding system of the room can ensure personal safety and normal operation of the instrument. For precision instruments that require uninterrupted power supply, an AC parameter regulated power supply is required to prevent harmonic interference.
 
3. Infrared carbon and sulfur analyzer working gas requirements: oxygen, purity> 99.5%
 
4. Installation tools